Basit öğe kaydını göster

dc.contributor.authorOnay, Aykut
dc.contributor.authorDogan, Uzeyir
dc.contributor.authorCiftci, Hakan
dc.contributor.authorCetin, Demet
dc.contributor.authorSuludere, Zekiye
dc.contributor.authorTamer, Ugur
dc.date.accessioned2020-06-25T18:29:20Z
dc.date.available2020-06-25T18:29:20Z
dc.date.issued2018
dc.identifier.citationclosedAccessen_US
dc.identifier.issn0947-7047
dc.identifier.issn1862-0760
dc.identifier.urihttps://doi.org/10.1007/s11581-018-2559-6
dc.identifier.urihttps://hdl.handle.net/20.500.12587/7242
dc.descriptionCetin, Demet/0000-0003-1186-4229en_US
dc.descriptionWOS: 000450111000030en_US
dc.description.abstractIn this report, an amperometric detection method for blood glucose level was developed benefiting from glucose oxidase (GO) enzyme immobilization on nanoparticle modified graphite rod (GR) electrode. Fe3O4-CS-Au magnetic nanoparticles were synthesized, characterized by TEM, UV-Vis, magnetometry, FTIR, and zeta potential measurements and used for the modification of GR electrode. This modified electrode was used for the detection of glucose level amperometrically at 0.6 V. The obtained calibration graph was linear in the range of 5-30mM glucose concentration with a coefficient of determination (R-2) 0.9971. The limit of detection (LOD) and limit of quantification (LOQ) values were calculated as 0.55 and 1.83mM, respectively. The modified GR electrode showed excellent selectivity in the presence of dopamine, ascorbic acid, and uric acid. The applicability of the developed method was examined in real blood samples by comparing the results obtained from commercial glucose sensor. This novel glucose detection method exhibited fast amperometric response, long storage time, and good selectivity.en_US
dc.language.isoengen_US
dc.publisherSpringer Heidelbergen_US
dc.relation.isversionof10.1007/s11581-018-2559-6en_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectGlucoseen_US
dc.subjectElectrochemical sensoren_US
dc.subjectMagnetic nanoparticleen_US
dc.subjectGraphite roden_US
dc.subjectChitosanen_US
dc.titleAmperometric glucose sensor based on the glucose oxidase enzyme immobilized on graphite rod electrode modified with Fe3O4-CS-Au magnetic nanoparticlesen_US
dc.typearticleen_US
dc.contributor.departmentKırıkkale Üniversitesien_US
dc.identifier.volume24en_US
dc.identifier.issue12en_US
dc.identifier.startpage4015en_US
dc.identifier.endpage4022en_US
dc.relation.journalIonicsen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US


Bu öğenin dosyaları:

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster