Pathogenic contamination of drinking water is critical in regard to human health. In this study, we investigated the electrical detection of single bacterium from drinking water. A microfluidics chip consisting of polydimethylsiloxane (PDMS) microchannel and gold microelectrodes was fabricated with conventional microfabrication techniques. Electrical characterizations were done with an LCR meter and the measurements were in good agreement with simulation results. The impact of channel and electrode dimensions was studied for the different type and size of particles, using both experimental and simulation techniques. In addition, the effect of excitation signal frequency and solution conductivity was analyzed employing both simulation and experimental methods. Finally, capacitive detection of a single Escherichia coli (E. coli) from drinking water was successfully carried out under optimum parameters and design geometries. (C) 2017 Elsevier B.V. All rights'reserved.
Açıklama
GULER, MUSTAFA TAHSIN/0000-0002-0478-3183
Anahtar Kelimeler
Microfluidics, Bacteria, Capacitive detection, Simulation
Kaynak
Sensors And Actuators A-Physical
WoS Q Değeri
Q2
Scopus Q Değeri
Q1
Cilt
269
Sayı
Künye
closedAccess