dc.description.abstractWithin the framework of quality management in medical services, the calibration gains great importance through appropriacy of medical devices. The aim of this study is to investigate the effects of temperature factor to the amount of gas output from the vaporizator and to enhance the consideration of these factors while adjusting the amount of gas to be given to the patient during the operation. In this study the performance of the anesthesia vaporizer (isoflurane and sevoflurane) was evaluated and it was investigated whether it the vaporizer supplied the dialed amount of anesthetic to the patient. Then, through changing the temperature of anesthetic device between 20-36 degrees C, the changes in the concentration of isoflurane and sevoflurane gases were measured. In conclusion, it was observed that there is no considerable change on the gas concentration in the 20-36 degrees C temperature range.en_US
dc.identifier.citationclosedAccessen_US
dc.identifier.endpage+en_US
dc.identifier.isbn978-1-4244-3605-7
dc.identifier.scopus2-s2.0-70350220569
dc.identifier.scopusqualityN/A
dc.identifier.startpage461en_US
dc.identifier.urihttps://hdl.handle.net/20.500.12587/4545
dc.identifier.wosWOS:000274345400120
dc.identifier.wosqualityN/A
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.language.isotr
dc.publisherIeeeen_US
dc.relation.ispartofBiyomut: 2009 14Th National Biomedical Engineering Meeting
dc.relation.publicationcategoryKonferans Öğesi - Uluslararası - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.titleThe Effects of Temperature Changes to the Anesthetic Gas Concentrationen_US
dc.typeConference Object

Dosyalar

Orijinal paket
Listeleniyor 1 - 1 / 1
[ X ]
İsim:
The Effects of Temperature Changes to the Anesthetic Gas Concentration.pdf
Boyut:
411.32 KB
Biçim:
Adobe Portable Document Format
Açıklama:
Tam Metin/Full Text